Skip to content

Public FDA record

Z-3088-2026: Velys Enabling Tech Robotic-Assist Station REF: 451611001

Recall record Z-3088-2026 for Medos International Sarl. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-3088-2026
Company in FDA record
Medos International Sarl
FDA classification date
Sep 2, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Software issue within robotic controller which may cause unintended movement of the robotic arm during bone preparation or instrument positioning.

Source facts

Action
On July 23, 2026, DePuy Synthes contacted affected consigenees via phone call concerning the recall. On July 24, 2026 Depuy issued a Urgent Medical Device Correction notification via sales associate visit. DePuy ask consignees to take the following actions: 1. Do not use the subject VELYS ROBOTIC-ASSIST STATION and quarantine it immediately. The VELYS. 2. Navigation Station and VELYS Camera Station can continue to be used following the Instructions for Use. 3. Complete, sign, and return the attached Business Response Form to OneMD-Field-Actions@its.jnj.com within three (3) business days of receipt of this notification. 4. Forward this notice to anyone in your facility that needs to be informed 5. Post a copy of this notice in a visible area for awareness and keep a copy for your records. 6. As with any medical device, adverse reactions or quality problems experienced with the use of this product should be reported to the FDA's MedWatch Adverse Event Reporting Program Online: https://www.fda.gov/safety/medwatch-fda-safety-information-and-adverse-event-reporting-program/reportingserious
Initiated
2026-07-23
Root cause
Under Investigation by firm
Distribution
US Nationwide distribution in the states of DE, FL, IL, KY, MD, MI, MN, NJ, NM, NY, OR, PA, TN, TX, VA, WA.
Recall number
Z-3088-2026
Recall status
Open, Classified
CDRH record ID
221615
FDA classified
2026-09-02
Recall event number
99467

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗