Z-3090-2024: paraPAC plus 310 ventilator kit with internal PEEP and CPAP, Item Number P310N

Recall record Z-3090-2024 for Smiths Medical ASD Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-3090-2024
Company in FDA record
Smiths Medical ASD Inc.
FDA classification date
Sep 20, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

Smiths Medical became aware of an issue related to a potential for the patient outlet connector to loosen/detach from the paraPAC Plus P300 and P310 ventilators impacting the active ventilation function.

Source facts

Action
Smiths Medical issued an URGENT MEDIC AL DEVICE CORRECTION notice to its consignees on05/31/2024 via letter. The notice explained the issue, potential risk, and requested the following: "Perform an inspection to determine if your devices are affected, per the instructions below: a. First perform a visual inspection to determine if the outlet connector is disconnected. b. Second evaluate the connector physically to determine if the outlet connector is loose or moves when placing a patient circuit on the connector or when removing it. c. If outlet connector remains tight after physical inspection, you can continue use of the device with heightened awareness and following all pre-use checks as per the user manual. d. If the outlet connector moves or feels loose, the device must be removed from use and repaired by Smiths Medical. Report the event to Global Complaint Management at globalcomplaints@icumed.com. Every use thereafter of every device, pre-use checks must be completed as described in the user manual and extra caution must be taken in inspecting the outlet connector prior to use and placing the patient circuit on the connector and during removal. Share this recall notification with all potential users of the devices to ensure they are aware of this recall and proposed mitigations." Those distributing the affected devices were directed to notify those to whom the devices were distributed. Device Correction Inquiries: contact - https://icumed.custhelp.com/app/marketaction
Initiated
2024-05-31
Root cause
Other
Distribution
Worldwide distribution.
Recall number
Z-3090-2024
Recall status
Open, Classified
CDRH record ID
208998
FDA classified
2024-09-20
Recall event number
94974

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗