Z-3098-2024: Spacelabs Ultraview DM3 Vital Signs Monitor (all variations -M, -MT, -N, -NT), Model Number 91330

Recall record Z-3098-2024 for Zoe Medical Incorporated. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-3098-2024
Company in FDA record
Zoe Medical Incorporated
FDA classification date
Sep 10, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

This product labeling does not adequately provide warning about the risks associated with the use of damaged batteries or batteries past the intended use life.

Source facts

Action
An URGENT MEDICAL DEVICE NOTIFICATION letter was sent to customers on 8/13/24. Actions being taken by Spacelabs Healthcare: The Operations Manual and Service Manual for the DM3 and DM4 have been updated to convey the risks associated with continuing to use the lithium batteries past the specified use life and to clarify that batteries older than two years, damaged, or that are no longer holding a charge must be replaced. Actions to be implemented by users: Users must download current versions of the Operations Manual and Service Manual and dispose of any previous copies. The updated product labeling is available via the Spacelabs user manual site at https://manuals.spacelabshealthcare.com/. Biomedical engineering must remove and replace any batteries that have been in use longer than two years, are damaged, or are no longer holding a charge. Contact Spacelabs Service to arrange for battery replacement. Use only manufacturer-supplied batteries. Biomedical engineering must arrange preventive maintenance to replace the batteries upon reaching the conclusion of their two-year use life. Acknowledgement Requested: Spacelabs requests that all customers acknowledge receipt and understanding of this notification by either mailing the included PDF copy of the response form or by completing the online form which can be accessed by using either the URL below or scanning the QR code: https://spacelabshealthcare.com/support/recall-reply-form/recall-reply-form-dm3-dm4-battery-life/ Please pass this notice on to those who need to be aware within your organization or to any organization where the potentially affected devices have been transferred. For additional information or technical assistance, please contact: Spacelabs Technical Support Spacelabs Healthcare, Inc 35301 SE Center Street Snoqualmie, WA 98065 1-800-522-7025 and select 2 for Technical Support
Initiated
2024-08-13
Root cause
Labeling design
Distribution
awaiting additional details
Recall number
Z-3098-2024
Recall status
Open, Classified
CDRH record ID
209630
FDA classified
2024-09-10
Recall event number
95160

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗