Due to a leak in the cooling system, fluids could potentially seep into the equipment cabinet. Loss of coolant can results in a cooling system failure and subsequent system failure. This leek can occur sporadically in affected systems.
Source facts
Action
Siemens mailed a Customer Safety Advisory Notice (CSAN) dated February 18, 2016, to affected customers informing them that their service organization will contact them to arrange a date to perform the corrective action (replacement of affected plastic couplings to prevent fluid leaks). Customers were advised to contact Siemens service organization for an appointment to schedule the work
For questions regarding this recall call 610-.448-6461.
Initiated
2016-02-18
Root cause
Unknown/Undetermined by firm
Terminated
2017-11-17
Distribution
Nationwide Distribution
Recall number
Z-3099-2017
Recall status
Terminated
CDRH record ID
158192
FDA record created
2017-08-29
Recall event number
77908