Open original FDA record ↗Recall record
- FDA identifier
- Z-3101-2017
- Company in FDA record
- Beckman Coulter Inc.
- FDA record creation date
- Sep 1, 2017
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The firm has identified that a fitting in some Slidemaker Stainer units may not completely seal after installation.
Source facts
- Action
- Beckman Coulter sent an Urgent Medical Device Recall letter dated June 20, 2017. The notification advised consignees to immediately contact a Beckman Coulter representative if a leak is observed to schedule a service, to inspect reagent supply drawer for spills, and to follow best laboratory practices when using the device (including Personal Protective Equipment). Beckman Coulter indicates that the issue will be resolved with the replacement of the correct fitting on all impacted instruments. Labs were advised to notify all staff of the recall and to retain the notification. They were also instructed to forward the recall notice to all customers, if any product was further distributed, and to complete and return the Customer Response form within ten days of receiving the recall notification. Completed forms are to be sent back to the firm via email to Regulatory.Notifications@beckmancoulter.com. Regulatory Affairs Mail Code 31-B06. Questions can be directed to beckmancoulter.com or 800-526-7694. Consignees outside of the US are advised to contact their Beckman Coulter representative in the case that questions do arise.
- Initiated
- 2017-06-20
- Root cause
- Employee error
- Terminated
- 2021-04-21
- Distribution
- Worldwide Distribution - USA (nationwide) Distribution to the states of : AL, AR, AZ, CA, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MO, MS, NC, NJ, NV, NY, OH, OR, PA, RI, SC, TN, TX, VA, WA, and WV., and to the countries of : Australia, Brazil, Canada, Chile, China, Czech Republic, Denmark, France, Greece, Hong Kong, India, Israel, Italy, Japan, Morocco, Portugal, South Korea, South Africa, Spain, Taiwan, Turkey, United Kingdom, Vietnam, and Switzerland.
- Recall number
- Z-3101-2017
- Recall status
- Terminated
- CDRH record ID
- 157861
- FDA record created
- 2017-09-01
- Recall event number
- 77778