Medical device for brachytherapy treatment may cause users to enter incorrect syntax and lead to potential misinterpretation of patient treatment plan.
Source facts
Action
Varian Medical Systems, Inc. sent a "Medical Device Recall" letter dated August 05, 2010 to all affected customers.
The letter included description of product affected and problem. Customers were advised to review template files to ensure that there are no characters immediately after the string in the customized dose quality parameters. Device users were requested to return a response form.
For information on this recall contact Varian Technical Support at (800) 360-7909.
Initiated
2010-08-11
Root cause
Software design
Terminated
2011-11-25
Distribution
Worldwide and Nationwide Distribution.
Recall number
Z-3103-2011
Recall status
Terminated
CDRH record ID
101796
FDA classified
2011-08-31
Recall event number
59217