Z-3104-2017: JTS External Drive Unit JTS Distal Femoral Extendible Implant Femur is intended for the replacement of d

Recall record Z-3104-2017 for Stanmore Implants Worldwide Ltd.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-3104-2017
Company in FDA record
Stanmore Implants Worldwide Ltd.
FDA record creation date
Sep 5, 2017
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Three (3) complaints were identified in which the potential for utilization of the JTS Grower Drive in the incorrect orientation in relation to the limb entry point of the device, together with the potential to select incorrect settings for the extension procedure. As a result, the prostheses in such cases failed to extend as intended.

Source facts

Action
Stanmore Implants sent an Urgent Field Safety Notice dated July 4, 2017, notifying customers of the recall. Additionally, the recalling firm requests that consignees immediately check their inventory for recalled product, quarantine any product on hand, inform SIW of any product on hand, complete the attached response form, and return the response for and affected product to the firm. It is requested that consignees distribute the notice internally to all affected parties and that they contact their local Stanmore representative to request training. If any adverse events are associated with this device, the firm asks that it be reported to them. For further questions, please call (972) 501-1402.
Initiated
2017-07-04
Root cause
Labeling design
Terminated
2021-02-16
Distribution
US Distribution to the states of : CA, FL, GA, KS, MI, NJ, OH, PA, SC, and TX.
Recall number
Z-3104-2017
Recall status
Terminated
CDRH record ID
157829
FDA record created
2017-09-05
Recall event number
77753

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗