Quick Disconnect Ratchet Handle may slip up the driver shaft or disengage leaving the shaft in the screw head.
Source facts
Action
Depuy Mitek issued Original Customer Notification Letter on 6/27/11 and an Updated letter to clarify the lot number information sent 6/29/11. Hospitals requested to immediately check all inventories for affected product and return for repalcement product. Also notify surgeons aware of this potential issue and potential risk of a delay in surgery. questions or concerns with regard to this program, please contact your local DePuy Mitek Representative or DePuy Mitek Customer Service directly at 1-800-382-4682, PROMPT 3.
Initiated
2011-06-27
Root cause
Process change control
Terminated
2012-01-19
Distribution
Nationwide distribution.
Recall number
Z-3106-2011
Recall status
Terminated
CDRH record ID
101878
FDA classified
2011-08-31
Recall event number
59229