Z-3106-2017: Spectranetics Bridge Occlusion Balloon catheter, Sheath Compatibility 12F, Guidewire Compatibility 0.035i

Recall record Z-3106-2017 for Spectranetics Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-3106-2017
Company in FDA record
Spectranetics Corporation
FDA classification date
Aug 26, 2017
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

Possible inability to pass the guidewire through the Bridge device, therefore, resulting in a potential delay of treatment.

Source facts

Action
The firm, Spectranetics, sent an "MEDICAL DEVICE RECALL" letter dated 08/07/2017 to customers via certified mail and an updated letter dated 8/17/2017 was sent on 8/23/2017. The letter described the product, problem and actions to be taken. The customers were Instructed to do the following: 1. For all Bridge inventory, we recommend physicians continue to follow product instructions to place a guide wire through the venous access site prior to the start of the procedure. Physicians may consider removing Bridge from packaging and placing the Bridge balloon over the wire prior to the start of the procedure to ensure a patent guidewire lumen if they believe the case is high risk. 2. All Bridge inventory will continue to have the potential for a guidewire lumen blockage until a solution is implemented. 3. Please bring your inventory into the case so that you have the back-up on hand, should you need it. 4. Always follow appropriate complication prevention and management protocols as it relates to patient preparation and surgical back up for lead extraction procedures. 5. Please complete the attached Acknowledgement and Receipt Form by EMAIL: Customer.Service@SPNC.com or FAX to: 1-877-447-2022. Your Spectranetics Sales Representative will be contacting you to facilitate the return and replacement of any remaining product in inventory once new inventory becomes available with an implemented fix; however, you may also reach Customer Service at 1-800-231-0978, Option 2. If you have additional questions please feel free to discuss with your local Spectranetics Sales Representative, or call the Vice President Quality Assurance directly at 719-447-2469. The Spectranetics Customer Service Department is also available to support you with any assistance you may need. Contact information:1-800-231-0978, Option 2 (Hours of Operation- Monday- Friday 7:00AM 5:00PM MST) or Email: Customer.service@spnc.com. An additional letter was sent via electronic notific
Initiated
2017-08-07
Root cause
Nonconforming Material/Component
Terminated
2020-04-14
Distribution
Worldwide Distribution: US (nationwide) and countries of: Australia, Austria, Belgium, Cyprus, Denmark, Finland, France, Germany, Hungary, Iceland, Ireland, Israel, Italy, Kuwait, Netherlands, Norway, Poland, South Africa, Sweden, Switzerland, and United Kingdom.
Recall number
Z-3106-2017
Recall status
Terminated
CDRH record ID
158269
FDA classified
2017-08-26
Recall event number
77934

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗