Open original FDA record ↗Recall record
- FDA identifier
- Z-3106-2026
- Company in FDA record
- Medical Action Industries, Inc. 306
- FDA classification date
- Sep 9, 2026
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class I
Reason stated in the source
Kits containing Sodium Chloride ampules subject to another recall are recalled due to potentially compromised sterility.
Source facts
- Action
- On 7/23/2026, "Urgent: Medical Device Recall" notices were emailed to customers who were asked to do the following: 1) Identify, Segregate, and Quarantine all impacted kits. 2) Add warning labels to affected kits to ensure all users are aware of the actions to take prior to use. A labeling template is enclosed for printing labels which can be affixed on all affected kits and cases in inventory. This label should be applied in a prominently visible location to end-users. The location chosen should not cover any other critical product information found on existing product labeling. If pre-printed labels are needed, please contact your sales representative. 3) REMOVE AND DISCARD the Sodium Chloride, 0.9%, USP 10 mL Flush ampules prior to use. 4) Customers (end-users): If you have received product from a distributor, please contact the distributor or your sales representative with any questions you may have. 5) Distributors: If any of the affected kits identified above were further distributed, please ensure all end-users are appropriately notified and maintain records of effectiveness. 6) Complete and return the Recall Response Form via email to GM-OMRA_RECALLS@owens-minor.com Report any quality problems or adverse events associated with the products listed in this notification to complaints@owens-minor.com If you have any questions, contact Regulatory Affairs Manager at email: GM-OMRA_RECALLS@owens-minor.com.
- Initiated
- 2026-07-23
- Root cause
- Nonconforming Material/Component
- Distribution
- US distribution to states of: IA, VA, MO, KS, FL, OH and LS.
- Recall number
- Z-3106-2026
- Recall status
- Open, Classified
- CDRH record ID
- 221992
- FDA classified
- 2026-09-09
- Recall event number
- 99613