Skip to content

Public FDA record

Z-3107-2026: Kit: Single Shot Epidural 20/Cs, REF: 59553; Kit: Single Shot Epidural 20/Cs, REF: 83945B. Containing Sod

Recall record Z-3107-2026 for Medical Action Industries, Inc. 306. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-3107-2026
Company in FDA record
Medical Action Industries, Inc. 306
FDA classification date
Sep 9, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

Kits containing Sodium Chloride ampules subject to another recall are recalled due to potentially compromised sterility.

Source facts

Action
On 7/23/2026, "Urgent: Medical Device Recall" notices were emailed to customers who were asked to do the following: 1) Identify, Segregate, and Quarantine all impacted kits. 2) Add warning labels to affected kits to ensure all users are aware of the actions to take prior to use. A labeling template is enclosed for printing labels which can be affixed on all affected kits and cases in inventory. This label should be applied in a prominently visible location to end-users. The location chosen should not cover any other critical product information found on existing product labeling. If pre-printed labels are needed, please contact your sales representative. 3) REMOVE AND DISCARD the Sodium Chloride, 0.9%, USP 10 mL Flush ampules prior to use. 4) Customers (end-users): If you have received product from a distributor, please contact the distributor or your sales representative with any questions you may have. 5) Distributors: If any of the affected kits identified above were further distributed, please ensure all end-users are appropriately notified and maintain records of effectiveness. 6) Complete and return the Recall Response Form via email to GM-OMRA_RECALLS@owens-minor.com Report any quality problems or adverse events associated with the products listed in this notification to complaints@owens-minor.com If you have any questions, contact Regulatory Affairs Manager at email: GM-OMRA_RECALLS@owens-minor.com.
Initiated
2026-07-23
Root cause
Nonconforming Material/Component
Distribution
US distribution to states of: IA, VA, MO, KS, FL, OH and LS.
Recall number
Z-3107-2026
Recall status
Open, Classified
CDRH record ID
221997
FDA classified
2026-09-09
Recall event number
99613

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.