Open original FDA record ↗Recall record
- FDA identifier
- Z-3108-2011
- Company in FDA record
- Accuray Incorporated
- FDA classification date
- Aug 31, 2011
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
As a result of an internal review, TomoTherapy has identified an issue with HiArt versions 4.0.x, and HD versions 1.0.x that we would like to bring to your attention. During DICOM export of plan level images with a non square exported Field of View (FOV), an anomaly in the process of squaring the plan level image may cause the image to shift with respect to ROIs and dose. When the anomaly occu
Source facts
- Action
- TomoTherapy sent a "Urgent Medical Device Correction" letter dated July 13, 2011, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. This issue will be remedied with HiArt software version 4.1 and HD software version 1.1 and later. If you have any questions, please contact the TomoTherapy Technical Solutions Center by e-mail "support@tomotherapy.com" . Please refer to the letter for what phone number you should call depending on where you are.
- Initiated
- 2011-07-13
- Root cause
- Software design
- Terminated
- 2014-03-18
- Distribution
- Worldwide Distribution-- USA (nationwide) including the states of AL, AK, AZ, AR, CA, CO, CT, DC, FL, GA, HI, IL, IN, IA, KS, KY, LA, MD, MA, MI, MN, MO, MT, NE, NJ, NM, NY, NC, ND, OH, OK, OR, PA, PR, SC, SD, TN, TX, UT, VA, WA, WV, WI and WY and the countries United Arab Emirates United Kingdom Turkey Switzerland Sweden Spain Saudi Arabia Poland Netherlands Italy Germany France Belgium Taiwan Singapore Malaysia Korea Japan India Hong Kong China Australia Mexico El Salvador and Canada
- Recall number
- Z-3108-2011
- Recall status
- Terminated
- CDRH record ID
- 102795
- FDA classified
- 2011-08-31
- Recall event number
- 59439