Open original FDA record ↗Recall record
- FDA identifier
- Z-3109-2017
- Company in FDA record
- Leonhard Lang Medizintechnik GmbH
- FDA classification date
- Sep 12, 2017
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class I
Reason stated in the source
Certain lots of the defibrillation electrodes model DF59N and DF59NC may have a delay in connecting or not connect at all with the defibrillator ZOLL AED PLUS. This may cause a situation in which a patient, who is in a life threatening condition and requires a defibrillation shock, cannot be treated in good time.
Source facts
- Action
- On approximately 08/08/2017, customers were notified via email of the recall. Instructions included to ensure all users and anyone concerned are aware of the recall, coordinate the replacement of affected product, once replacement electrodes are received document destruction of affected product on-site, and notify customers if the product had been further distributed.
- Initiated
- 2017-08-08
- Root cause
- Component design/selection
- Terminated
- 2020-06-09
- Distribution
- Worldwide distribution. US distribution in the states of : FL & TX. Austria, China, France, Germany, Greece, Hungary, India, Italy, Panama, Philippines, Poland, Saudi Arabia, Serbia, South Africa, South Korea, Spain, United Arab Emirates, United Kingdom.
- Recall number
- Z-3109-2017
- Recall status
- Terminated
- CDRH record ID
- 158310
- FDA classified
- 2017-09-12
- Recall event number
- 77949