Skip to content

Public FDA record

Z-3109-2026: Halyard IV Start Kit Non-Sterile REF: KRIV26-02

Recall record Z-3109-2026 for AVID Medical, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-3109-2026
Company in FDA record
AVID Medical, Inc.
FDA classification date
Sep 4, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Convenience kits contain BD ChloraPrep Clear - 1 mL and FREPP Clear 1.5 mL Applicators which lacks sterility assurance due to to wrinkles in the paper lidding which may breach the seal area.

Source facts

Action
On July 14, 2026, Avid Medical Inc. a subsidiary of Owens Minor issued an Urgent Medical Device Recall to affected consignees via email. I Avid Medical Inc. requested consignees to take the following actions: 1. Identify, segregate, and quarantine all impacted product identified. 1a. Discontinue use of the BD ChloraPrep" Clear - 1 mL and FREPP" Clear 1.5 mL Applicators. 1b. Add warning labels to affected AVID kits to ensure all users are aware of the actions to discard the impacted component prior to use. A labeling template is enclosed for printing labels which can be affixed on all affected kits and cases in inventory. This label should be applied in a prominently visible location to end-users. The location chosen should not cover any other critical product information found on existing product labeling. 2. Customers (end-users): If you received product from a distributor, please contact the distributor or your sales representative with any questions you may have. 3. Distributors: If any of the affected kits identified above were further distributed, please ensure all end-users are appropriately notified and maintain records of effectiveness. 4. Complete the attached Recall Response Form and return by July 31, 2026.
Initiated
2026-07-14
Root cause
Packaging process control
Distribution
US Nationwide distribution in the state of NE.
Recall number
Z-3109-2026
Recall status
Open, Classified
CDRH record ID
221740
FDA classified
2026-09-04
Recall event number
99525

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗