Z-3110-2017: MOSAIQ Medical Charged-Particle Radiation Therapy System

Recall record Z-3110-2017 for Elekta, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-3110-2017
Company in FDA record
Elekta, Inc.
FDA record creation date
Sep 6, 2017
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The table shift in the vertical direction was inverted during the use of CMA (Couch Move Assistant). Therapy may be directed to an incorrect area.

Source facts

Action
Consignees were to receive a recall notice by mail dated 06/30/2017. The notice identifies the reason for recall and goes on to instruct consignees to run verification tests on units to ensure they are operating properly. Only applicable tests should be performed on the unit. If the unit passes verification tests, then the attached acknowledgment form is to be signed and returned to Elekta. If the unit fails verification, consignees are instructed to contact their local Elekta Care Support Center for assistance. If a unit is no longer in use, consignees are asked to sign and return the acknowledgement form, noting that the unit is no longer in use. The firm asks that the recall notice be placed in an accessible area to users until the field action is closed. Appropriate personnel should also be provided the content of the recall notification.
Initiated
2017-06-30
Root cause
Software in the Use Environment
Terminated
2020-10-26
Distribution
Worldwide distribution. US nationwide to AK, AL, AR, AZ, CA, CT, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, WY, District of Columbia, Hawaii and Puerto Rico. ALBANIA, ALGERIA, ANTIGUA AND BARBUDA, ARGENTINA, AUSTRALIA, AUSTRIA, BAHRAIN, BANGLADESH, BELARUS, BELGIUM, BENIN, BOTSWANA, BRAZIL, BULGARIA, CANADA, CHILE, CHINA, COLOMBIA, CROATIA, CUBA, CYPRUS, CZECH REPUBLIC, DENMARK, ECUADOR, EGYPT, FINLAND, FRANCE, GABON, GEORGIA, GERMANY, GHANA, GREECE, HONDURRAS, HONG KONG, HUNGARY, INDIA, INDONESIA, IRAN, IRAQ, IRELAND, ISRAEL, ITALY, JAPAN, JORDAN, KAZAKHSTAN, KENYA, LAOS, LIBYA, LITHUANIA, MALAYSIA, MALTA, MEXICO, MOROCCO, MYANMAR, NAMIBIA, NETHERLANDS, NEW ZEALAND, NIGERIA, NORWAY, PAKISTAN, PANAMA, PERU, PHILIPPINES, POLAND, PORTUGAL, ROMANIA, RUSSIA, SAUDI ARABIA, SERBIA, SINGAPORE, SLOVAKIA, SLOVENIA, SOUTH AFRICA, SOUTH KOREA, SPAIN, SRI LANKA, SWEDEN, SWITZERLAND, TAIWAN, THAILAND, TUNISIA, TURKEY, UNITED KINGDOM, UKRAINE, VENEZUELA, VIETNAM and ZIMBABWE
Recall number
Z-3110-2017
Recall status
Terminated
CDRH record ID
157834
FDA record created
2017-09-06
Recall event number
77758

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗