Utilizing higher fluences at 1064nm wavelength with current cryogen parameters may result in a spark/loud noise during treatment, potentially resulting in patient burn.
Source facts
Action
Firm began notifying customers on 8/6/26.
Customers were instructed to discontinue use of previous parameters and utilize the new revised clinical cryogen parameters. Updated cryogen parameters were provided with the letter.
If you have questions or need assistance, please contact Clinical Support at 844-365-5060 or via email at clinical@cynosurelutronic.com.
Initiated
2026-08-06
Root cause
Device Design
Distribution
Worldwide distribution - US Nationwide and the countries of Canada and European Union.
Recall number
Z-3110-2026
Recall status
Open, Classified
CDRH record ID
221941
FDA classified
2026-09-04
Recall event number
99601