Z-3115-2017: Spectra-System/Legacy Guided Surgery Handle Kit, Device Listing Number D245883. The Guided Handle Kit (G

Recall record Z-3115-2017 for Implant Direct Sybron Manufacturing, LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-3115-2017
Company in FDA record
Implant Direct Sybron Manufacturing, LLC
FDA record creation date
Sep 11, 2017
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The inside diameter of the 2.3mm small drill guide inserts have a smaller diameter than the specifications require.

Source facts

Action
Implant Direct sent an Safety Alert /Advisory Notice letter dated June 26, 2017. Instructions included to review and remove inventory for return to Implant Direct, complete and return the acknowledgement and recall return form, and contact customers and notify the of the recall if product was further distributed. For further questions, please call (888) 649-6425.
Initiated
2017-06-26
Root cause
Process control
Terminated
2021-06-16
Distribution
Worldwide Distribution - USA (nationwide) Distribution to the states of : AZ, CA, CO, CT, FL, ID, IL, KS, MI, MO, MS, NC, NE, NH, NJ, NM, NV, NY, OH, OR, PA, SC, TN, TX, UT, VA, WA., and to the countries of : Canada, Netherlands, Spain, Great Britain, Germany and Switzerland.
Recall number
Z-3115-2017
Recall status
Terminated
CDRH record ID
156495
FDA record created
2017-09-11
Recall event number
77554

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗