Skip to content

Public FDA record

Z-3120-2017: Cell Marque, Podoplanin (D2-40 Mab), 7 mL, REF PA0796, IVD, The antibody is intended for in vitro diagn

Recall record Z-3120-2017 for Leica Biosystems Richmond Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-3120-2017
Company in FDA record
Leica Biosystems Richmond Inc.
FDA record creation date
Sep 12, 2017
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

A delay in transit may have contributed to nonconformity as the product was stored outside of defined shipping and handling temperature requirements.

Source facts

Action
Advance Cancer Diagnostics sent an Urgent Medical Device Recall letter dated July 12, 2017, to all affected customers. The letter identified the product, the problem and the action to be taken by the customer. Customers were instructed to identify any affected products, discard them, and complete and return the Medical Device Acknowledgement Form by fax to 1-847-236-3000. For questions regarding this recall call 815-678-2000.
Initiated
2017-07-14
Root cause
Storage
Terminated
2024-08-02
Distribution
Nationwide Distribution to AR, AZ, CA, CT, FL, GA, IL, IN, KY, LA, MI, MN, NC, NH, NV, NY, OR, PA, SC, TN, TX, VA, and WI.
Recall number
Z-3120-2017
Recall status
Terminated
CDRH record ID
158567
FDA record created
2017-09-12
Recall event number
78055

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗