Z-3120-2026: HeartSine SAM 350P 350P US Device, US English - 2 Min CPR UDI-DI codes: 0506016712067 350P US Device, US

Recall record Z-3120-2026 for HeartSine Technologies Ltd. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-3120-2026
Company in FDA record
HeartSine Technologies Ltd
FDA classification date
Sep 4, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Due to a damaged (over stretched) membrane tail that causes an intermittent fault.

Source facts

Action
On 07/28/2026, several versions of an "Urgent: Medical device recall" letter that outlines multiple issues: 1) Stryker identified an issue with the membrane tail, an internal component, which could cause an intermittent fault. This failure could occur at any point when the device is holding a charge in preparation to deliver therapy, while delivering a shock, or after shock delivery. The device becomes inoperable after the failure occurs. 2) Stryker has also identified that a limited number of HeartSine samaritan PAD devices were distributed without the required maintenance insert. This insert contains periodic maintenance instructions implemented as an additional risk control measure. While most devices contain the maintenance insert, this voluntary correction is being initiated to ensure all end users have access to the supplemental maintenance information. This issue was identified through a customer complaint. For the membrane tail issue (issue 1), Customers are instructed to:
Initiated
2026-07-27
Root cause
Process control
Distribution
Worldwide - U.S. Nationwide distribution including in the states of AK, AL, AZ, CA, CT, DC, FL, GA, HI, IA, ID, IL, IN, KS, KY, MA, MD, ME MI, MN, NC, NJ, NM, NV, NY, OH, OK, OR, PA, SD,TN, TX, UT, VA, WA, WI, WV and WY. The countries of Argentina, Australia, Brazil, Canada, Chile, Colombia, Hong Kong, Indonesia, ISRAEL, JAPAN, Malaysia, Netherlands, Philippines, POLAND, SINGAPORE, South Korea, Taiwan, Thailand and UNITED KINGDOM.
Recall number
Z-3120-2026
Recall status
Open, Classified
CDRH record ID
221890
FDA classified
2026-09-04
Recall event number
99568

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗