Open original FDA record ↗Recall record
- FDA identifier
- Z-3124-2024
- Company in FDA record
- Fresenius Medical Care Holdings, Inc.
- FDA classification date
- Sep 11, 2024
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Machines equipped with BPM3 Blood Pressure Modules (Blood pressure monitor TM-2917B with firmware version S005) will not record Blood Pressure measurements and does not alarm if the cuff tubing is kinked close to the cuff.
Source facts
- Action
- An IMPORTANT MEDICAL DEVICE CORRECTION notification letter dated 8/8/24 was sent to customers. Facilities and Programs with the 2008T Hemodialysis Machine with Blood Pressure Module BPM3 should: "Visually inspect the blood pressure tubing for kinking during each monitoring and safety check. Correct any kinks if observed. "Verify visually that each scheduled blood pressure measurement was completed successfully each time period as per clinic procedures and state regulations. For further information please reference the 2008T Hemodialysis System Operator's Manual (P/N 490122 Rev. AA) https://fmcna.com/support/product-support-documents/operators-manuals-hemodialysis-hd/ Fresenius Medical Care is working diligently on a correction and will communicate when updated firmware is available. Please provide this Customer Notice to all those who need to be aware within your organization. Please complete and return the attached reply form. On 12/19/2024, Fresenius Medical sent a follow up notification letter indicating there is a new BPM3 Blood Pressure Module (firmware version S007) available that fixes this issue. Instructions included how to determine if you have affected BPM3 firmware and if it is affected, how to return the unit and order the replacement part. Customers were also instructed to notify personnel and customers, and to complete and return the response form.
- Initiated
- 2024-08-08
- Root cause
- Under Investigation by firm
- Distribution
- US Nationwide. Canada.
- Recall number
- Z-3124-2024
- Recall status
- Open, Classified
- CDRH record ID
- 209659
- FDA classified
- 2024-09-11
- Recall event number