Z-3124-2024: Blood Pressure Module BPM3 TM-2917B (Spare Part Numbers 191378 or 191389) with firmware version S005 inst

Recall record Z-3124-2024 for Fresenius Medical Care Holdings, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-3124-2024
Company in FDA record
Fresenius Medical Care Holdings, Inc.
FDA classification date
Sep 11, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Machines equipped with BPM3 Blood Pressure Modules (Blood pressure monitor TM-2917B with firmware version S005) will not record Blood Pressure measurements and does not alarm if the cuff tubing is kinked close to the cuff.

Source facts

Action
An IMPORTANT MEDICAL DEVICE CORRECTION notification letter dated 8/8/24 was sent to customers. Facilities and Programs with the 2008T Hemodialysis Machine with Blood Pressure Module BPM3 should: "Visually inspect the blood pressure tubing for kinking during each monitoring and safety check. Correct any kinks if observed. "Verify visually that each scheduled blood pressure measurement was completed successfully each time period as per clinic procedures and state regulations. For further information please reference the 2008T Hemodialysis System Operator's Manual (P/N 490122 Rev. AA) https://fmcna.com/support/product-support-documents/operators-manuals-hemodialysis-hd/ Fresenius Medical Care is working diligently on a correction and will communicate when updated firmware is available. Please provide this Customer Notice to all those who need to be aware within your organization. Please complete and return the attached reply form. On 12/19/2024, Fresenius Medical sent a follow up notification letter indicating there is a new BPM3 Blood Pressure Module (firmware version S007) available that fixes this issue. Instructions included how to determine if you have affected BPM3 firmware and if it is affected, how to return the unit and order the replacement part. Customers were also instructed to notify personnel and customers, and to complete and return the response form.
Initiated
2024-08-08
Root cause
Under Investigation by firm
Distribution
US Nationwide. Canada.
Recall number
Z-3124-2024
Recall status
Open, Classified
CDRH record ID
209659
FDA classified
2024-09-11
Recall event number
95169

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗