During positioning within the system's 3D to 3D matching mode while transferring CBCT images, when the Float Image Selection button is pressed, it may display CBCT images that are insufficient in number.
Source facts
Action
Between 08/05-09/2024, the firm communicated/hand-delivered an "URGENT: MEDICAL DEVICE RECALL" letter to customers informing them that during positioning in the Positioning Image Analysis Systems (PIAS) 3D3D matching mode, while transferring Cone Beam CT systems (CBCT) images from CT Image Reconstruction Software (CIRS) to PIAS, when the Float Image Selection button on the PIAS screen is pressed, it may display CBCT images that are insufficient in number. This may reveal a clear displacement in the Y-direction (body axis direction) compared to the planning CT. There may potentially be a possibility of completing 3D3D patient positioning approximately 5mm off in the body axis direction (with 5 images missing) and proceeding with treatment.
Customers are instructed, until the software has been replaced with the corrected version, all users of the device are directed to follow the interim countermeasure.
When users operate as described in the operation manual, no such event should occur.
Countermeasure references:
Engineering Document:
PBT-40E-1042 Instruction for patient positioning assistance using CBCT
[PROBEAT-FR] Operation Manual TS.16F0147-1
Page.1-105 / 1-108
[PROBEAT-CR] Operation Manual TS.16F0147-1
Page. 1-4-20 / 1-4-23
[PROBEAT-V] Operation Manual TS.16A1850-1E
Page.1-126 / 1-131
For questions/assistance -
Initiated
2024-05-27
Root cause
Software design
Distribution
U.S. Nationwide distribution in the states of TN, TX, and WA.
Recall number
Z-3125-2024
Recall status
Open, Classified
CDRH record ID
208834
FDA classified
2024-09-12
Recall event number
94928