Open original FDA record ↗Recall record
- FDA identifier
- Z-3128-2011
- Company in FDA record
- Beckman Coulter Inc.
- FDA classification date
- Sep 2, 2011
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
The recall was initiated because Beckman Coulter has confirmed the microbial contamination in the conjugate diluent bottles included in the Manual Inhibin A Elisa kit (Part Number DSL-10-28100-4), kit lot number 110761.
Source facts
- Action
- Beckman Coulter initiated an "URGENT PRODUCT CORRECTION" on June 15, 2011, via a telephone script to the only one customer impacted by the recalled product. The telephone script provided an explanation of the problem identified and the actions to be taken. Beckman Coulter recommended that the customer discontinue use of the recalled product as a preventative measure. Beckman will send a representative to the customer's facility to replace the recalled product in all remaining kits of the affected lot. The customer was instructed to share this information with laboratory staff. For any questions concerning this information, contact Beckman Coulter at 952-368-7874.
- Initiated
- 2010-11-10
- Root cause
- Other
- Terminated
- 2012-07-11
- Distribution
- (USA) Nationwide Distribution in the state of CA
- Recall number
- Z-3128-2011
- Recall status
- Terminated
- CDRH record ID
- 103283
- FDA classified
- 2011-09-02
- Recall event number
- 59685