Z-3128-2026: StatStrip Glucose Hospital Meter System. Model No: 63685. Glucose test system.

Recall record Z-3128-2026 for Nova Biomedical Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-3128-2026
Company in FDA record
Nova Biomedical Corporation
FDA classification date
Sep 11, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

A software synchronization/timing issue in the barcode scanning and patient data processing component that causes the device to intermittently retain a previously scanned barcode including patient identification (ID) barcodes and quality control (QC) solution ID barcodes when a new barcode is scanned, resulting in the prior barcode information remaining displayed on the screen for the current test. When this defect occurs and the operator accepts the incorrectly displayed patient information without recognizing the mismatch, the current patient's glucose result will be incorrectly assigned to the previous patient's medical record and transmitted to the data management system (DMS), and the current patient's medical record will not receive a result, leaving their glucose status unrecorded and their clinical management without the information needed to guide treatment decisions. The previous patient, whose medical record now contains a glucose result that does not belong to them, may receive inappropriate clinical intervention based on that erroneous result, particularly in critically ill patients, neonates, and infants who cannot communicate symptoms or self-treat. The defect also affects QC barcode scanning: if a QC scan triggers the same retention behavior, QC results may be attributed to an incorrect identifier or QC tests may fail without the user recognizing the underlying cause. This defect does not affect the analytical accuracy of glucose measurements.

Source facts

Action
Nova Biomedical issued an "URGENT MEDICAL DEVICE RECALL" Customer Advisor Notice to its consignees on 7/29/2026 via email. The notice explained the issue, potential risk, and requested consignees follow the below specific recommendations until a permanent resolution can be provided. 1. Verifying patient information displayed on the meter exactly matches the scanned patient identification after every patient barcode scan and before accepting the patient's record or result. 2. If the displayed patient information does not match the scanned patient identification, discontinue testing with the meter and notify your administrator. 3. Nova recommends that a copy of this notice is provided to all end users within your facilities that use the StatStrip (Gen 2) Glucose Hospital Meter Systems for patient testing and care. Additionally, Nova recommends placing a copy of this notice in all areas where the meters are stored and used. 4. Complete and return the acknowledgement form to appropriate emails: North America- Novameters@novabio.com; European Union- EU-QA@novabio.com and all other international customers-Intlservice@novabio.com. Nova is actively working on a planned firmware update to address this issue. As a precautionary measure, Nova has implemented a shipment hold on all StatStrip (Gen 2) Glucose Hospital Meter Systems while corrective actions are being developed and validated. Efforts are underway for the development and validation of the firmware update to address this issue. Nova will communicate availability as soon as validation activities are completed targeting September 2026. If you have questions regarding this notice, please contact Nova Biomedical Technical Support at 1 (800) 545-6682 (USA) or at (800) 263-5999 (Canada). Nova Biomedical Technical Support can be reached 24 hours a day, 7 days a week, 365 days a year. For customers outside of the USA or Canada, please contact your authorized Nova Biomedical distributor.
Initiated
2026-07-29
Root cause
Software design
Distribution
Worldwide distribution: US (nationwide) and OUS (foreign) countries of: Austria, Belgium, Bolivia, Brazil, Canada, Cayman Islands, Chile, Croatia, Dominican Rep., Finland, France, Germany, Ireland, Israel, Italy, Netherlands, New Zealand, South Korea, Switzerland, Thailand, Turkey, United Kingdom.
Recall number
Z-3128-2026
Recall status
Open, Classified
CDRH record ID
221819
FDA classified
2026-09-11
Recall event number
99546

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗