Z-3134-2017: VITEK(R) 2 GP, REF 21342, IVD, UDI 03573026131920, bioMerieux, Inc. VITEK(R) 2 is an automated system con

Recall record Z-3134-2017 for Biomerieux Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-3134-2017
Company in FDA record
Biomerieux Inc
FDA record creation date
Sep 18, 2017
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Customer reports have indicated occurrences of atypical negative O129R reactions for Enterococcus casseliflavus ATCC(R) 700327(TM) in association with VITEK(R) 2 GP ID Lot 2420240403 and 2420303403, causing failure of the Quality Control (QC) testing for the VITEK(R) 2 GP ID card. The expected O129R reaction for ATCC 700327(TM) strain is positive.

Source facts

Action
bioMerieux sent an Urgent Product Removval Notice to all affected consignees (Subsidiaries) of bioMrieux, impacted by the identified issue, were notified of the FSCA and corrective action (FSCA notification) electronically on August 25, 2017. . The directions to the user are as follows: "Please implement the following actions at this time: We request you take the following action at this time: " Please confirm this letter has been distributed and reviewed by all appropriate personnel within your organization. " Please refrain from using the referenced test kits (VITEK 2 GP ID, Lots 2420240403 and 2420303403) in your laboratory. Destroy any of this product in your inventory, and contact your local bioMrieux representative for credit or replacement. " Please store this letter with your bioMrieux system documentation. " Complete the attached Acknowledgement Form and return it to your local bioMrieux representative." Consignees with questions were instructed to contact their local representative.
Initiated
2017-08-25
Root cause
Under Investigation by firm
Terminated
2020-11-09
Distribution
US and Angola, Argentina, Austria, Belarus, Belgium, Burkina Faso, Cameroon, Canada, China, Congo, Croatia, Egypt, France, Germany, Greece, Guam, Hungary, Israel, Italy, Latvia, Luxembourg, Macedonia, Mali, Mexico, Morocco, Netherlands, Nicaragua, Oman, Panama, Poland, Saudi Arabia, Serbia, South Africa, South Korea, Spain, Swaziland, Switzerland, Tunisia, Turkey, Turkmenistan, United Arab Emirates, United Kingdom
Recall number
Z-3134-2017
Recall status
Terminated
CDRH record ID
158270
FDA record created
2017-09-18
Recall event number
77935

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗