Open original FDA record ↗Recall record
- FDA identifier
- Z-3140-2026
- Company in FDA record
- CooperSurgical, Inc.
- FDA classification date
- Sep 18, 2026
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class I
Reason stated in the source
Potential to exhibit a loose connection between the flexible tubing and the nasal cannula, during assembly or while in use, which may result in loss of CPAP pressure, possibly reducing the effectiveness of treatment and lead to decreased oxygenation until connection is restored.
Source facts
- Action
- CooperSurgical notified consignees on about 08/11/2026 via "URGENT Medical Device Recall" letter. Consignees were notified of the issue and potential hazard, and were instructed to identify affected units on hand, if any loose connection is recognized during the use of the CPAP, replace the tubing with another tubing or CPAP device, and remove all product of the subject lots from inventory and note the number of units to be returned on the attached Recall Response Form that is provided. Distributors were instructed to remove all product of the subject lots from inventory and note the number of units to be returned on the attached Distributor Only Acknowledgement Form. Additionally, they were instructed to communicate the recall notification to all customers that they distributed these devices to.
- Initiated
- 2026-08-11
- Root cause
- Device Design
- Distribution
- Nationwide distribution to CT, FL, GA, IA, IL, IN, MD, MI, MS, NJ, NY, OH, PA, VA. International distribution to Brazil and Peru.
- Recall number
- Z-3140-2026
- Recall status
- Open, Classified
- CDRH record ID
- 222063
- FDA classified
- 2026-09-18
- Recall event number
- 99643