Z-3141-2017: Spine Referencing Instrumentation, Double Spinous Process Clamp, Tall When used with a Medtronic StealthS

Recall record Z-3141-2017 for Medtronic Navigation, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-3141-2017
Company in FDA record
Medtronic Navigation, Inc.
FDA record creation date
Sep 21, 2017
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Potential for Spinal Clamps to be damaged when opened beyond intended limits, either no opening once attached to a patient's spinous process or leaving a component behind in the patient's body following breakage.

Source facts

Action
Medtronic sent an Urgent Medical Device Correction letter dated June 27, 2017, to all affected consignees. The letter notifies consignees of the product correction and requests that practitioners follow the attached use instructions for devices. Practitioners are to identify product damage before each procedure by opening and closing the device jaw with the T-handle. It is instructed that practitioners not continue turning the screw if resistance is met, as this may break the washer. Clamps are not to be used if the jaws remain closed when attempting to open a clamp or if the clamp screw back is out of the clamp. It is asked that this check is done to all inventory and any damaged product is to be reported to Medtronic at 1-800-595-9709 and quarantined. Revised product is available. Consignees are asked to complete and return the provided Consignee Response Form by email to the firm at RS.NavFCA@medtronic.com or by fax to 651-367-7075. Any questions can be directed to 1-800-595-9709.
Initiated
2017-06-29
Root cause
Device Design
Terminated
2023-07-24
Distribution
Worldwide Distribution - US including AL, AR, AZ, CA, CO, CT, DC, FL, GA, IA, AD, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, SC, SD, TN, TX, UT, VA, WA, WI, WV, & WY. Foreign: Albania, Australia, Bangladesh, Belgium, Bulgaria, Canada, Croatia, Czech Republic, Denmark, Finland, France, Germany, Greece, India, Indonesia, Israel, Italy, Japan, Kenya, Latvia, Luxembourg, Malaysia, Martinique, Netherlands, New Zealand, Philippines, Poland, Portugal, Romania, Russia, Singapore, Spain, Sweden, Switzerland, Thailand, Turkey, United Kingdom, & Vietnam.
Recall number
Z-3141-2017
Recall status
Terminated
CDRH record ID
158157
FDA record created
2017-09-21
Recall event number
77894

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗