Z-3142-2026: Philips PIC iX Essentials REF:867093

Recall record Z-3142-2026 for Philips North America. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-3142-2026
Company in FDA record
Philips North America
FDA classification date
Sep 11, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Due to software malfunction, patient information system may freeze and become unresponsive.

Source facts

Action
On July 31, 2026, Philips North America LLC issued an urgent medical device correction to affected consignees via USPS certified mail. Philips requested consignees to take the following actions: Users: 1. In cases where central monitoring has halted, please ensure that triaging of patients at bedside and a change in workflow occurs to more closely monitor patient status. Ensure audible alarm volume is on high in-patient rooms in case bedside devices disconnect and do not provide visual indicators that communication with the central. 2. Pass this notice to those who need to be aware within your organization or to any organization where affected product(s) have been potentially transferred. 3. Place this Urgent notification with the documentation of the Philips Patient Information Center (PIC) iX and associate devices in a place where it is most likely to be seen and viewed. 4. Implement the Technical Solution established by Philips as soon as available within the defined timeframe. Distributors: 1. Send the attached URGENT Medical Device Correction to each customer to whom you have distributed any affected product(s) 2. Perform a good faith effort to reach each customer a minimum of three (3) times, and if possible, using multiple contact methods. 3. Implement the Technical Solution established by Philips as soon as available within the timeframe communicated by Philips, and confirm with Philips the implementation of the correction of all affected product(s). In case any product(s) cannot be corrected, inform Philips of the reason why the Technical Solution could not be implemented.
Initiated
2026-07-31
Root cause
Software design
Distribution
Worldwide - US Nationwide distribution including in the states and territories of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY, PR and the countries of Andorra, Utd. Arab. Emir., Argentina, Austria, Australia, Aruba, Bosnia-Herz., Bangladesh, Belgium, Bulgaria, Bahrain, Bolivia, Brazil, Canada, Switzerland, Chile, Colombia, Costa Rica, Cyprus, Czech Republic, Germany, Denmark, Dominican Rep, Algeria, Estonia, Spain, Finland, Fiji, Faroe Islands, France, United Kingdom, Gibraltar, Guadeloupe, Greece, Hong Kong, Croatia, Hungary, Indonesia, Ireland, Israel, India, Iraq, Iceland, Italy, Jordan, Japan, Kenya, South Korea, Kuwait, Lebanon , Sri Lanka, Lithuania, Luxembourg, Latvia, Libya, Morocco, Monaco, Montenegro, Macedonia, Macao, Martinique, Maldives, Mexico, Malaysia, Namibia, Nicaragua, Netherlands, Norway, New Zealand, Oman, Panama, Peru, Philippines, Pakistan, Poland, Puerto Rico, Portugal, Qatar R¿union, Romania, Serbia, Russian Fed., Saudi Arabia, Sweden, Singapore, Slovenia, Slovakia, Thailand, Turkmenistan, Tunisia, T¿rkiye, Taiwan, Ukraine, Uzbekistan, Vietnam, South Africa.
Recall number
Z-3142-2026
Recall status
Open, Classified
CDRH record ID
221926
FDA classified
2026-09-11
Recall event number
99566

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗