Open original FDA record ↗Recall record
- FDA identifier
- Z-3145-2024
- Company in FDA record
- GE Healthcare GmbH
- FDA classification date
- Sep 13, 2024
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The National Institute of Standards and Technology (NIST) has identified in the National Vulnerability Database (NVD) that NexGen Healthcare Mirth Connect, prior to version 4.4.1, has a potential vulnerability allowing the deserialization of untrusted data (CVE-2023-43208). Mirth products may have been installed and configured with your ViewPoint / ViewPoint 6 system by GE HealthCare. This vulnerability could allow a malicious actor to access the system and potentially manipulate patient data.
Source facts
- Action
- GE Healthcare issued an URGENT MEDICAL DEVICE CORRECTION notice on 08/08/2024 via US Mail. The notice explained the issue, potential risk, and requested the following: You can continue to use your device by following the instructions below: 1.Ensure ViewPoint / ViewPoint 6 is operating in a secure hospital Network such that: a. only authorized personnel have access, b. protections against denial of service and wiretapping/network sniffing are installed by the network operators, and c. access from the Internet is limited where possible 2.Ensure ViewPoint / ViewPoint 6 is operating in a secure Windows environment including confirmation that all relevant software is up to date and operating systems updates are installed. Please ensure all potential users in your facility are made aware of this safety notification.
- Initiated
- 2024-08-08
- Root cause
- Device Design
- Distribution
- US Nationwide distribution including in the states of Alabama, Alaska, Arizona, Arkansas, California, Colorado, Connecticut, Delaware, District of Columbia, Florida, Georgia, Idaho, Illinois, Indiana, Iowa, Kansas, Kentucky, Louisiana, Maine, Maryland, Massachusetts, Michigan, Minnesota, Mississippi, Missouri, Montana, Nebraska, Nevada, New Hampshire, New Jersey, New Mexico, New York, North Carolina, North Dakota, Ohio, Oklahoma, Oregon, Pennsylvania, Rhode Island, South Carolina, South Dakota, Tennessee, Texas, Utah, Vermont, Virginia, Washington, West Virginia, Wisconsin.
- Recall number
- Z-3145-2024
- Recall status
- Open, Classified
- CDRH record ID
- 209643
- FDA classified
- 2024-09-13
- Recall event number
- 95167