Z-3148-2026: Automated Impella Controller (AIC) with the below product descriptions and corresponding Product Codes. 1

Recall record Z-3148-2026 for Abiomed, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-3148-2026
Company in FDA record
Abiomed, Inc.
FDA classification date
Sep 16, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

The potential for purge cassette recognition issues due to the failure of the purge flag component of the purge pressure sensor assembly.

Source facts

Action
On August 12, 2026 URGENT MEDICAL DEVICE RECALL (REMOVAL) letters were sent to customers. Actions to be Taken: 1. Hospital inventory can continue to be used while awaiting service from Abiomed to ensure continuity of care. 2. Abiomed s field service team will contact you to schedule the specified replacement. 3. In the event of these purge flag component failures and associated alarms, as per IFU, first reinsert purge disc to address the alarm. If that action does not clear the alarm, switch to a backup Automated Impella Controller. 4. Ensure cleaning of the AIC housing with mild detergent as specified in the IFU. 5. Ensure future routine preventive maintenance is performed annually on your AIC(s) following the guidance in the IFU 6. Forward this notice to anyone in your facility that needs to be informed. 7. If any of the subject products have been forwarded to another facility, contact that facility and provide them with this notice. 8. Post a copy of this notice in a visible area for awareness.
Initiated
2026-08-12
Root cause
Under Investigation by firm
Distribution
Domestic: Nationwide Distribution. International: United Arab Emirates, Austria, Australia, Belgium, Brunei Darussalam, Brazil, Canada, Switzerland, Chile, China, Colombia, Costa Rica, Cyprus, Czechia, Germany, Denmark, Spain, Finland, France, United Kingdom, Greece, Hong Kong, Croatia, Hungary, Ireland, Israel, India, Italy, Japan, Kuwait, Luxembourg, Mexico, Malaysia, Netherlands, Norway, New Zealand, Panama, Pakistan, Poland, Serbia, Saudi Arabia, Sweden, Singapore, Slovenia, Thailand, Taiwan;
Recall number
Z-3148-2026
Recall status
Open, Classified
CDRH record ID
222148
FDA classified
2026-09-16
Recall event number
99671

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗