Skip to content

Public FDA record

Z-3152-2024: 2090 Programmers with serial number prefixes PKK0 and PKK1: a) 2090, Product Description: PRGM 2090 P

Recall record Z-3152-2024 for Medtronic Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-3152-2024
Company in FDA record
Medtronic Inc.
FDA classification date
Sep 13, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Medtronic CareLink 2090 programmers with serial number prefixes PKK0 and PKK1 have the potential for autonomous cursor motion when Finger Touchscreen capability is enabled by software version 3.2 or higher.

Source facts

Action
Medtronic issued a MEDICAL DEVICE CORRECTION notice to its consignees on 07/11/2024 via email or UPS 2-day delivery. The notice explained the issue, potential risk, and provided the following: "Instrument Management Recommendations Software updates are necessary to maintain proper programmer function. Medtronic representatives will assist in performing the software update on all Medtronic CareLink 2090 programmers and assess proper function. Medtronic representatives will assist with returning programmers needing repair or replacement. " Following review of this letter, please complete the enclosed Confirmation Form and email it to RS.FCACRM@medtronic.com to confirm receipt of this notification. " Please forward this notice to others who need to be aware within your organization."
Initiated
2024-07-11
Root cause
Device Design
Distribution
Worldwide distribution.
Recall number
Z-3152-2024
Recall status
Open, Classified
CDRH record ID
209676
FDA classified
2024-09-13
Recall event number
95176

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗