Open original FDA record ↗Recall record
- FDA identifier
- Z-3152-2024
- Company in FDA record
- Medtronic Inc.
- FDA classification date
- Sep 13, 2024
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Medtronic CareLink 2090 programmers with serial number prefixes PKK0 and PKK1 have the potential for autonomous cursor motion when Finger Touchscreen capability is enabled by software version 3.2 or higher.
Source facts
- Action
- Medtronic issued a MEDICAL DEVICE CORRECTION notice to its consignees on 07/11/2024 via email or UPS 2-day delivery. The notice explained the issue, potential risk, and provided the following: "Instrument Management Recommendations Software updates are necessary to maintain proper programmer function. Medtronic representatives will assist in performing the software update on all Medtronic CareLink 2090 programmers and assess proper function. Medtronic representatives will assist with returning programmers needing repair or replacement. " Following review of this letter, please complete the enclosed Confirmation Form and email it to RS.FCACRM@medtronic.com to confirm receipt of this notification. " Please forward this notice to others who need to be aware within your organization."
- Initiated
- 2024-07-11
- Root cause
- Device Design
- Distribution
- Worldwide distribution.
- Recall number
- Z-3152-2024
- Recall status
- Open, Classified
- CDRH record ID
- 209676
- FDA classified
- 2024-09-13
- Recall event number
- 95176