Z-3153-2024: Ivenix Infusion System (IIS), LVP Software; LVP-SW-0004, Software Version 5.9.1 and prior. Infusion pump

Recall record Z-3153-2024 for Fresenius Kabi USA, LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-3153-2024
Company in FDA record
Fresenius Kabi USA, LLC
FDA classification date
Sep 26, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

The software has anomalies that have the potential to cause alarms, nonfunctioning pump, or unresponsive screen while continuing therapy. These could cause serious patient harm or death.

Source facts

Action
An URGENT Medical Device Field Correction Important Safety Update notification letter dated 8/30/24 was sent to customers. Required Actions for Users: 1. Please fill out the Customer Reply Form attached to this letter to acknowledge receipt. 2. If your facility further distributes or transfers products to satellite sites or other locations, please ask these locations to contact the email address on the Customer Reply Form so we can send notice to those satellites directly. 3. Install new software version (5.9.2) for the Ivenix Large Volume Pump (LVP). See details below. How To Obtain and Install Software Version 5.9.2 To request installation, please reach out promptly to your Fresenius Kabi representative (1-855- 354-6387 or Ivenix_support@fresenius-kabi.com). They will push the new software update to your pumps. When the LVP version 5.9.2 software update has been sent to the LVP, the following prompt appears before the LVP is shut down. Customer support - email Ivenix_su pport@fresenius-kabi.com Customer support - phone 1-855-354-6387 Normal business Hours are Monday through Friday 8:30 AM to 5 PM Eastern Time, After Hours Supported with Voicemail Messaging
Initiated
2024-08-30
Root cause
Software design
Distribution
US Nationwide.
Recall number
Z-3153-2024
Recall status
Open, Classified
CDRH record ID
209801
FDA classified
2024-09-26
Recall event number
95240

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗