Open original FDA record ↗Recall record
- FDA identifier
- Z-3156-2024
- Company in FDA record
- Mazor Robotics Ltd
- FDA classification date
- Sep 16, 2024
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Software anomalies which can result in two separate issues: A lagging screen display during a navigated procedure; or the potential for a mismatch between the navigated screw and the screw information presented on the screen.
Source facts
- Action
- Medtronic/Mazor Robotics, Ltd. issued an URGENT: Medical Device Software Correction to its consignees on 07/17/2024 via 2-day tracked mailing. The notice explained the issue, potential risk, and provided a workaround until the software update is installed. Medtronic will schedule a time to perform a software update that will resolve these anomalies. Medtronic field service engineers will update all Mazor X systems from version 5.1.1 to version 5.1.2 or later. Questions: contact Medtronic Technical Services at 1-888-826-5603 or rs.navtechsupport@medtronic.com .
- Initiated
- 2024-07-17
- Root cause
- Device Design
- Distribution
- US Nationwide distribution in PR.
- Recall number
- Z-3156-2024
- Recall status
- Open, Classified
- CDRH record ID
- 209698
- FDA classified
- 2024-09-16
- Recall event number
- 95189