Z-3156-2024: Mazor X. Model Number REF TPL0059; stereotaxic orthopedic instrument

Recall record Z-3156-2024 for Mazor Robotics Ltd. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-3156-2024
Company in FDA record
Mazor Robotics Ltd
FDA classification date
Sep 16, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Software anomalies which can result in two separate issues: A lagging screen display during a navigated procedure; or the potential for a mismatch between the navigated screw and the screw information presented on the screen.

Source facts

Action
Medtronic/Mazor Robotics, Ltd. issued an URGENT: Medical Device Software Correction to its consignees on 07/17/2024 via 2-day tracked mailing. The notice explained the issue, potential risk, and provided a workaround until the software update is installed. Medtronic will schedule a time to perform a software update that will resolve these anomalies. Medtronic field service engineers will update all Mazor X systems from version 5.1.1 to version 5.1.2 or later. Questions: contact Medtronic Technical Services at 1-888-826-5603 or rs.navtechsupport@medtronic.com .
Initiated
2024-07-17
Root cause
Device Design
Distribution
US Nationwide distribution in PR.
Recall number
Z-3156-2024
Recall status
Open, Classified
CDRH record ID
209698
FDA classified
2024-09-16
Recall event number
95189

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗