When the key switch is turned on, the exposure conditions are set to the default values (85kV, 1.6 mAs, approx. 0.13mGy@ SID 1m) intended for adult chest imaging. In addition, if exposure conditions are set from the console while the key switch is off, the set exposure conditions will not be reflected.
Source facts
Action
Customer and Dealer notification will occur by means of letters alerting customers and dealers to the potential for overexposure to children if the conditions are not checked after the key switch is turned on. The notices also include the plan for the affected mobile x-ray systems. The system will be upgraded, free of charge, with updated software so that the exposure values set by the user are reflected regardless of the state of the key switch.
Initiated
2023-11-23
Root cause
Radiation Control for Health and Safety Act
Distribution
US Nationwide and Worldwide
Recall number
Z-3158-2024
Recall status
Open, Classified
CDRH record ID
209977
FDA classified
2024-09-25
Recall event number
95320