Skip to content

Public FDA record

Z-3160-2011: Picis ED PulseCheck-Software, transmission & storage, patient data Service Pack v5.0 L and M This produ

Recall record Z-3160-2011 for Picis Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-3160-2011
Company in FDA record
Picis Inc.
FDA classification date
Sep 7, 2011
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 17, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Software does not display appropriate allergy interaction warning

Source facts

Action
Picis issued notification on 08/01/11 to affected clients, explaining the potential for the error, with instructions to monitor documented patient allergies closely when prescribing medication orders in ED PulseCheck, using Muitum drug database content. Picis will track the progress on acknowledgement of notification letter receipt through the implementation of software upgrade, for all potentially affected Customers. For questions regarding this recall call 781-557-3000.
Initiated
2011-08-01
Root cause
Software design
Terminated
2013-06-27
Distribution
Nationwide Distribution
Recall number
Z-3160-2011
Recall status
Terminated
CDRH record ID
103528
FDA classified
2011-09-07
Recall event number
59737

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗