Z-3160-2024: GE Healthcare Centricity Universal Viewer Zero Footprint Client, Model Numbers: a) 2110344-008; b) 21

Recall record Z-3160-2024 for GE Healthcare. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-3160-2024
Company in FDA record
GE Healthcare
FDA classification date
Sep 17, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Footprint Client (ZFP) versions v6.0 SP9.x and SP10.x where the latest addended report is not shown by default to the user. The issue occurs when the addendum is created on the same day as the original report but at a different time, and only when launched on the following browsers. " Internet Explorer 10 (IE10) " Internet Explorer 11 (IE11) " Firefox.

Source facts

Action
GE Healthcare issued an URGENT MEDICAL DEVICE CORRECITON notice to its consignees on 07/23/2023 via letter. The notice explained the issue, potential risk to the patient, and provided the following actions be taken: "You can continue to use your device. Please ensure all potential users in your facility are made aware of this safety notification and the recommended actions below. Follow one of the below recommendations to ensure the latest addendum is being reviewed: 1. Use Microsoft Edge or Google Chrome browsers when launching ZFP Viewer. OR 2. If you cannot use the browsers listed in #1, to find the addendum / latest report, review all the reports available for the study in the Series Selector. Navigation to the series selection can be found in the User Manual under Section 2.3.3 of The Series Selector, with subheading as Open a report/document using series selector"." GE HealthCare will correct all affected products at no cost to you. A GE HealthCare representative will contact you to arrange for the correction. For questions or concerns regarding this notification, please contact GE HealthCare Service at 1-800-437-1171 or your local Service Representative.
Initiated
2024-07-23
Root cause
Software Design Change
Distribution
Worldwide distribution.
Recall number
Z-3160-2024
Recall status
Open, Classified
CDRH record ID
209798
FDA classified
2024-09-17
Recall event number
95238

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗