Skip to content

Public FDA record

Z-3162-2011: Ratcheting Torque Limiting Handle, Catalog Number 52-1012, Orthofix Spinal Implants, Firebird Spinal Fixa

Recall record Z-3162-2011 for Orthofix, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-3162-2011
Company in FDA record
Orthofix, Inc.
FDA classification date
Sep 7, 2011
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 17, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Upon reaching torque limit, there is a possibility the product may jam or exhibit a ratchet direction switch from forward to reverse.

Source facts

Action
Orthofix, Inc. sent an "URGENT PRODUCT RECALL NOTIFICATION" letter dated April 1, 2011 to all affected customers. The letter describes the product, problem, and the actions to be taken by the customers. The letter instructs customers to identify and remove the affected product from use. Customers were advised to use the combination of the Ratching R Handle and the Set Screw Driver if they experience problems with the Ratching Torque Limiting Handle. A Tracking and Verification form was attached for customers to complete and return to the Orthofix Inc. Regulatory Affairs Department via fax at 214-937-3322. Contact your Account Services Representative at 888-298-5700 for questions regarding the removal and return of the affected product.
Initiated
2011-04-01
Root cause
Device Design
Terminated
2012-03-23
Distribution
Worldwide Distribution-USA (nationwide) and Puerto Rico and the countries of Australia and Germany.
Recall number
Z-3162-2011
Recall status
Terminated
CDRH record ID
99424
FDA classified
2011-09-07
Recall event number
58473

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗