Open original FDA record ↗Recall record
- FDA identifier
- Z-3162-2011
- Company in FDA record
- Orthofix, Inc.
- FDA classification date
- Sep 7, 2011
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Upon reaching torque limit, there is a possibility the product may jam or exhibit a ratchet direction switch from forward to reverse.
Source facts
- Action
- Orthofix, Inc. sent an "URGENT PRODUCT RECALL NOTIFICATION" letter dated April 1, 2011 to all affected customers. The letter describes the product, problem, and the actions to be taken by the customers. The letter instructs customers to identify and remove the affected product from use. Customers were advised to use the combination of the Ratching R Handle and the Set Screw Driver if they experience problems with the Ratching Torque Limiting Handle. A Tracking and Verification form was attached for customers to complete and return to the Orthofix Inc. Regulatory Affairs Department via fax at 214-937-3322. Contact your Account Services Representative at 888-298-5700 for questions regarding the removal and return of the affected product.
- Initiated
- 2011-04-01
- Root cause
- Device Design
- Terminated
- 2012-03-23
- Distribution
- Worldwide Distribution-USA (nationwide) and Puerto Rico and the countries of Australia and Germany.
- Recall number
- Z-3162-2011
- Recall status
- Terminated
- CDRH record ID
- 99424
- FDA classified
- 2011-09-07
- Recall event number
- 58473