Open original FDA record ↗Recall record
- FDA identifier
- Z-3164-2026
- Company in FDA record
- Nuclein LLC
- FDA classification date
- Sep 14, 2026
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Due to a software issue
Source facts
- Action
- On 10/22/2025 a "Customer Advisory Notification" via email to customers informing them of a software update to deal with potential of aborted test runs due to Error Code 0x25000012 (Pressure wet filter error). Customer were informed that an over-the-air software update will automatically be available when the DASH Instrument is connected to the internet. For questions or assistance - Nuclein Customer Support at 1-888-992-DASH (3274) or customersupport@nuclein.com On 07/30/2026, an updated "URGENT: MEDICAL DEVICE CORRECTION" Letter was email to customers. The updated letter informed customer of the potential issues associated with the software issue that was not addressed in the initial letter sent on 10/22/2025.
- Initiated
- 2025-10-22
- Root cause
- Process control
- Distribution
- US Nationwide distribution in the states of Alabama, Delaware, Florida, Georgia, Kansas, Massachusetts, , Maryland, Maine, Michigan, North Carolina, New York, Ohio, South Carolina, Tennessee, Texas, Virginia, Vermont.
- Recall number
- Z-3164-2026
- Recall status
- Open, Classified
- CDRH record ID
- 221403
- FDA classified
- 2026-09-14
- Recall event number
- 99356