Skip to content

Public FDA record

Z-3173-2026: GM Helix Implant, REF: 109.984;

Recall record Z-3173-2026 for Straumann USA LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-3173-2026
Company in FDA record
Straumann USA LLC
FDA classification date
Sep 15, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

There is a discrepancy between the product identification on the packaging and the dimensions of the implant contained inside.

Source facts

Action
On August 12, 2026, Urgent Field Safety Action letters were sent to customers. Actions to be taken: 1. Identify and segregate any units in your stock with carton packaging labeled as Article 109.984 Helix GM Implant, 4.0 x 11.5 mm, L ot VZPF4 2. If the article is still in your inventory, then stop use/ distribution of the product immediately and quarantine / segregate physically. 3. Please return the affected items to the attention of Todd Shevlin at Straumann USA, LLC for replacement as indicated on the Customer Confirmation Form. 4. If the product has already been implanted, implant removal is not required, and no additional patient follow -up is required solely due to this recall 5. In all cases, complete and return the enclosed Customer Confirmation Form with the affected product using the return UPS label provided to: Straumann USA, LLC. If product is not being returned, the completed Customer Confirmation Form can be scanned and emailed to todd.shevlin@straumann.com
Initiated
2026-08-12
Root cause
Under Investigation by firm
Distribution
US Nationwide distribution in the states of AL, AZ, CA, CO, FL, GA, IL, KS, LA, MA, MI, MO, NC, NJ, NV, NY, OH, OK, OR, PA, SC, TX, VA, WI;
Recall number
Z-3173-2026
Recall status
Open, Classified
CDRH record ID
222258
FDA classified
2026-09-15
Recall event number
99707

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗