Routine internal review of Device History Records identified six VITEK(R) 2 AST card lots were manufactured with less drug (gentamicin) than required for the Gentamicin High-Level Resistance (ghlr01n) screen test due to an error when making 1500mL and 2000mL media well preparations (solutions).
Source facts
Action
The firm initiated the recall by email on 08/13/2018. The firm requested the consignee discard the impacted product and contact the local bioMerieux representative for credit.
Initiated
2018-08-13
Root cause
Process control
Terminated
2020-11-02
Distribution
Worldwide Distribution US Nationwide and the countries of: Turkey, Austria, Switzerland, Czech Republic, Germany, Netherlands,
Recall number
Z-3179-2018
Recall status
Terminated
CDRH record ID
166752
FDA record created
2018-09-15
Recall event number
80860