Certain 2860 IsoFlex LAL support surfaces were assembled with a non-conforming fire barrier.
Source facts
Action
Urgent Medical Device Correction notification letters dated 8/13/18 were distributed to customers. The letters included the following instructions:
1. Locate the units listed on the attached business reply form and identify the
address where they can be serviced.
2. Return the enclosed business reply form to confirm receipt of this notification by
fax 269 488 8691 or email productfieldaction@stryker.com.
3. Upon receipt of the completed business reply form, a member of Strykers field
service personnel will contact you to arrange for the replacement of the fire
barrier on your affected IsoFlex LAL units.
4. If you have questions regarding the continued safe use of your products, please
contact Technical Support by calling +1 800 327 0770 and selecting opt 2.
5. If you have loaned or sold any of the products listed in this letter, please forward
a copy of this notice to the new users and advise us of their new location in the
space provided on the business reply form.
6. If you have disposed of any of these units and they are no longer in use, please,
advise us of their obsolescence by providing us with their serial numbers in the
space provided on the business reply form.
Initiated
2018-08-13
Root cause
Mixed-up of materials/components
Terminated
2020-05-21
Distribution
US Nationwide in the states of: KY, NH, OH, PA, and TX.
Recall number
Z-3181-2018
Recall status
Terminated
CDRH record ID
166820
FDA record created
2018-09-15
Recall event number
80894