Skip to content

Public FDA record

Z-3181-2026: Blood Culture Collection Venipuncture Kit 20/Cs, REF: 77451; Blood Culture Kit 20/Cs, REF: 81834B; Blood

Recall record Z-3181-2026 for Medical Action Industries, Inc. 306. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-3181-2026
Company in FDA record
Medical Action Industries, Inc. 306
FDA classification date
Sep 16, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Blood culture convenience kits contain a component subject to a drug recall that due to an an open or incomplete seal on the packaging of the applicator may compromise the component's sterile barrier and if used may lead to localized skin or soft tissue infection.

Source facts

Action
On 7/23/2026, correction notices were emailed to customers who were asked to do the following: 1) Identify, segregate, and quarantine all impacted product. 2) Discontinue use of the BD ChloraPrep Clear - 1 mL Applicators 3) Add warning labels to affected MAI kits to ensure all users are aware of the actions to discard the impacted component prior to use. A labeling template is enclosed for printing labels which can be affixed on all affected kits and cases in inventory. This label should be applied in a prominently visible location to end-users. The location chosen should not cover any other critical product information found on existing product labeling. 4) Customers (end-users): If you received product from a distributor, please contact the distributor or your sales representative with any questions you may have. 5) Distributors: If any of the affected kits identified above were further distributed, please ensure all end-users are appropriately notified and maintain records of effectiveness. 6) Complete and return Recall Response Form via email to GM-OMRA_RECALLS@owens-minor.com Report any quality problems or adverse events associated with the products listed in this notification to complaints@owens-minor.com.
Initiated
2026-07-23
Root cause
Material/Component Contamination
Distribution
US Nationwide distribution in the states of OH, IN, NC.
Recall number
Z-3181-2026
Recall status
Open, Classified
CDRH record ID
222021
FDA classified
2026-09-16
Recall event number
99628

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗