Open original FDA record ↗Recall record
- FDA identifier
- Z-3181-2026
- Company in FDA record
- Medical Action Industries, Inc. 306
- FDA classification date
- Sep 16, 2026
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Blood culture convenience kits contain a component subject to a drug recall that due to an an open or incomplete seal on the packaging of the applicator may compromise the component's sterile barrier and if used may lead to localized skin or soft tissue infection.
Source facts
- Action
- On 7/23/2026, correction notices were emailed to customers who were asked to do the following: 1) Identify, segregate, and quarantine all impacted product. 2) Discontinue use of the BD ChloraPrep Clear - 1 mL Applicators 3) Add warning labels to affected MAI kits to ensure all users are aware of the actions to discard the impacted component prior to use. A labeling template is enclosed for printing labels which can be affixed on all affected kits and cases in inventory. This label should be applied in a prominently visible location to end-users. The location chosen should not cover any other critical product information found on existing product labeling. 4) Customers (end-users): If you received product from a distributor, please contact the distributor or your sales representative with any questions you may have. 5) Distributors: If any of the affected kits identified above were further distributed, please ensure all end-users are appropriately notified and maintain records of effectiveness. 6) Complete and return Recall Response Form via email to GM-OMRA_RECALLS@owens-minor.com Report any quality problems or adverse events associated with the products listed in this notification to complaints@owens-minor.com.
- Initiated
- 2026-07-23
- Root cause
- Material/Component Contamination
- Distribution
- US Nationwide distribution in the states of OH, IN, NC.
- Recall number
- Z-3181-2026
- Recall status
- Open, Classified
- CDRH record ID
- 222021
- FDA classified
- 2026-09-16
- Recall event number
- 99628