Open original FDA record ↗Recall record
- FDA identifier
- Z-3184-2011
- Company in FDA record
- Straumann Manufacturing Inc.
- FDA classification date
- Sep 13, 2011
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Software version 8.0 has a software anomaly, in which the length is not adjusted accordingly when it is changed after the surgical sleeve position planning step has been completed.
Source facts
- Action
- Staumann USA notified accounts by an Urgent Field Safety Notice letter dated August 4, 2011 and delivered via Federal Express Delivery Service. The letter identified the affected product and explained the reason for the recall. Customers were asked to review the letter, take appropriate actions, including completing and faxing the enclosed. Customer Confirmation Form. Straumann released a corrected software upgrade (coDiagnostiX 8.0.2.3672), which is available for download when users go online and activate current software. If the user does not have internet access a replacement CD can be ordered from Technical Support at 866 - 531 -7365. Questions should be directed to Technical Support Monday through Friday from 8:00 am to 8:00 pm Eastern Time.
- Initiated
- 2011-08-04
- Root cause
- Software design
- Terminated
- 2012-05-23
- Distribution
- Nationwide Distribution
- Recall number
- Z-3184-2011
- Recall status
- Terminated
- CDRH record ID
- 103755
- FDA classified
- 2011-09-13
- Recall event number
- 59797