Z-3185-2018: PATHFAST Myoglobin II reagent cartridge Catalog Number: PF1111-K PATHFAST NTproBNP is a product for in-v

Recall record Z-3185-2018 for Polymedco, Inc.,/Polymedco Cancer Diagnostic Products, Llc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-3185-2018
Company in FDA record
Polymedco, Inc.,/Polymedco Cancer Diagnostic Products, Llc
FDA record creation date
Sep 17, 2018
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

A pinhole caused by the laser printer in the aluminum sheet over the ALP reagent wells caused leakage through the pinhole. If the cartridge with leakage is used, higher measurement value as low as +12% and no higher than +20% bias may be obtained.

Source facts

Action
The firm, Polymedco, initiated a recall by telephone and followed with email on 4/30/18 to inform users to inspect the reagent cartridges for evidence of leakage from the ALP well and to discard cartridges that had evidence of leakage. Polymedco will replace any affected reagent cartridges. If you have any questions, contact Polymedco Technical Support at the 1.800.431.2123 Ext 285.
Initiated
2018-04-30
Root cause
Process control
Terminated
2019-04-17
Distribution
US Distribution to states of: AR, CA, FL, ID, LA, MN, NY and TX.
Recall number
Z-3185-2018
Recall status
Terminated
CDRH record ID
166226
FDA record created
2018-09-17
Recall event number
80724

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗