Z-3188-2026: Boston Scientific Corporation StoneSmart Connect Console Contents: (1) Power Cord, (1) Power Adaptor, (1)

Recall record Z-3188-2026 for Boston Scientific Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-3188-2026
Company in FDA record
Boston Scientific Corporation
FDA classification date
Sep 16, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

A subset of power adapters used with the connect console may exhibit higher-than-expected resistance in the electrical grounding path. As a result, the power adaptor may fail electrical safety testing during installation.

Source facts

Action
On July 22, 2026, Boston Scientific Corporation issued a medical device correction to affected consignees via FedEx letter, Boston Scientific Corporation requested consignees to take the following actions: 1. Continue to use the affected devices according to the Instructions for Use. 2. Forward this notice to relevant personnel and any facilities to which the impacted device was transferred. 3. Complete and return the enclosed Acknowledgement Form. 4. Begin using the replacement power adaptor upon receipt and dispose of the affected power adaptor in accordance with local requirements.
Initiated
2026-07-22
Root cause
Nonconforming Material/Component
Distribution
Worldwide - US Nationwide distribution including in the states of AZ, FL, CA, OR, IL, NC, NE, UT, CO, OH, AR, NM, AL, WI, TN, IN, GA, OH, WI, KY, CT, KS, SC, IL, NE, MO, MS, DE, TX, NY, WV, MA, MT, WA, WY, NV, HI, ID, MO, ME, WV, NE, OK, MN, LA, WV, NV, PA, NH, NJ, MI, MO, AK, VA, MS, WY, ME and the countries of Australia, Finland, France, Germany, Great Britain, Greece, Ireland, Italy, Lebanon, Norway, Poland, Portugal, Saudi Arabia, Slovakia, Spain, Sweden, Switzerland, Turkiye, Canada, Hong Kong, Singapore, Taiwan, Argentina, Chile, Japan.
Recall number
Z-3188-2026
Recall status
Open, Classified
CDRH record ID
222047
FDA classified
2026-09-16
Recall event number
99630

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗