Open original FDA record ↗Recall record
- FDA identifier
- Z-3190-2018
- Company in FDA record
- Siemens Medical Solutions USA, Inc
- FDA record creation date
- Sep 18, 2018
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Temperature of the detector may fall out of its designed working range (cool down) due to incorrectly configured temperature control unit. This unintentional system behavior may occur while the unit is powered down for long periods (e.g. at night).
Source facts
- Action
- Urgent Medical Device Correction notification letters dated 7/27/18 were distributed to customers. The letters informed the customers of the following planned actions: Our service organization will contact you shortly to arrange a date to perform this corrective action with Update Instruction AX001/18/S. Please feel free to contact our service organization for an earlier appointment at 1-800-888-7436. Following the corrective action, the cause is eliminated and a recurrence of the fault is prevented. Additionally, Siemens is developing a solution to restore the system back to the original settings. This solution will be distributed to all affected customers upon completion.
- Initiated
- 2018-07-29
- Root cause
- Device Design
- Terminated
- 2019-05-17
- Distribution
- US nationwide distribution.
- Recall number
- Z-3190-2018
- Recall status
- Terminated
- CDRH record ID
- 166917
- FDA record created
- 2018-09-18
- Recall event number
- 80905