Skip to content

Public FDA record

Z-3190-2018: ARTIS Pheno Interventional Fluoroscopic X-Ray System, Model Number 10849000

Recall record Z-3190-2018 for Siemens Medical Solutions USA, Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-3190-2018
Company in FDA record
Siemens Medical Solutions USA, Inc
FDA record creation date
Sep 18, 2018
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Temperature of the detector may fall out of its designed working range (cool down) due to incorrectly configured temperature control unit. This unintentional system behavior may occur while the unit is powered down for long periods (e.g. at night).

Source facts

Action
Urgent Medical Device Correction notification letters dated 7/27/18 were distributed to customers. The letters informed the customers of the following planned actions: Our service organization will contact you shortly to arrange a date to perform this corrective action with Update Instruction AX001/18/S. Please feel free to contact our service organization for an earlier appointment at 1-800-888-7436. Following the corrective action, the cause is eliminated and a recurrence of the fault is prevented. Additionally, Siemens is developing a solution to restore the system back to the original settings. This solution will be distributed to all affected customers upon completion.
Initiated
2018-07-29
Root cause
Device Design
Terminated
2019-05-17
Distribution
US nationwide distribution.
Recall number
Z-3190-2018
Recall status
Terminated
CDRH record ID
166917
FDA record created
2018-09-18
Recall event number
80905

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗