Z-3191-2018: ddR Formula B X-ray System, Model ddR Formula B X-ray system used for imaging

Recall record Z-3191-2018 for Swissray Medical. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-3191-2018
Company in FDA record
Swissray Medical
FDA classification date
Sep 18, 2018
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Potential for bucky (the part that holds the grid and is moveable to position the patient) moving unexpectedly.

Source facts

Action
The firm, Swissray Medical, issued official communication to customers as a Action Notification Report/work Instructions on November 18 2016. The Work Instructions state the affected devices, the problem symptoms, causes of the problem, remedy, any tools required, and the step-by-step corrective actions. The manufacturer, Swissray Medical AG, distributed the Work Instructions for the FCO and the Action Notification Report Form to all distributors and service entities who would be performing corrective actions in the field. The Swissray Field Engineer then presented the Report to the customer (end user facility) and communicated with the customer in-person while performing the corrective action on-site. When corrective actions were completed, both the customer and service personnel then sign the Report to confirm that the customer was informed about the corrective action and that corrective action was completed. The Report Form states the title and unique designation of the corrective action, whether it was mandatory by the manufacturer or was an optional product improvement, and the unique document ID number of the Work Instructions for the FCO. The customers were instructed to Complete the FCO Action Notification Report and send to: Swissray Medical AG expert Center Turbistrasse 25-27 CH-6280 Hochdorf Switzerland Or scan and email to service@swissray.com with FCO in the subject line. If you have any questions, contact QA/RA Coordinator at 908-372-6408 or email: Deniz.Kortan@swissray.com.
Initiated
2016-11-18
Root cause
Device Design
Distribution
Worldwide Distribution: US (Nationwide) and to countries of: Austria, Dubai, China. Germany, Libya, Malaysia, Russia, and South Africa.
Recall number
Z-3191-2018
Recall status
Open, Classified
CDRH record ID
167170
FDA classified
2018-09-18
Recall event number
80981

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗