Z-3193-2018: VITROS Chemistry Products CREA Slides, Product Code 680 2584 Product Usage: VITROS CREA Slides (Product

Recall record Z-3193-2018 for Ortho-Clinical Diagnostics. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-3193-2018
Company in FDA record
Ortho-Clinical Diagnostics
FDA record creation date
Sep 19, 2018
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class III

Reason stated in the source

The slides did not meet the current claims for Limit of Blank (LoB), Limit of Detection (LoD) and Limit of Quantification (LoQ) for serum and urine samples and therefore affecting the low end of the measuring range for this assay. The current low end measuring range is 0.05mg/dL for serum and 1.2mg/dL for urine. Based on the CREA testing results for the LoQ, the revised low end CREA measuring range is 0.15mg/dL for serum and 3.2mg/dL for urine.

Source facts

Action
On 7/11/18, Important Product Correction Notification letters were distributed to customers. The letters instructed customers to perform the following actions: VITROS Cl- Slides and VITROS CREA Slides Instructions for Use will be updated to contain the revised information described in this letter; we will notify you upon availability. In the interim, retain a copy of this customer letter for your records. Calibration Diskette and ADD Data Release Version (DRV) 5992 (and above) will support the revised measuring range for VITROS CREA Slides. We anticipate that DRV 5992 will be available beginning on July 28, 2018 by eConnectivity or will be sent via your normal Cal Disk or ADD mailing. Prior to the availability of the revised IFU, be aware that if a serum or plasma sample contains a Triglyceride concentration of 600 mg/dL or higher, the Chloride result may be positively biased by approximately 2.1 mmol/L. Follow your normal laboratory procedures as you would for other known assay interferences. Note: It is acceptable to continue using VITROS Cl- Slides. Upon availability, install the Calibration Diskette or ADD to implement the revised Measuring Range: For VITROS 4600, 5600 or 5,1 FS Systems: Install ADD DRV 5992 or above, Select All Assay Data option For VITROS 250/350 Systems: Install Calibration Diskette DRV 5992 or above Refer to the enclosure for additional instructions to update the Measuring Range. Update laboratory procedures and your Laboratory Information System (LIS), as necessary. Retain this letter until we send a notification to alert you when the revised IFUs are available. Complete the Confirmation of Receipt form and return by July 25, 2018. Post this notification by each system that processes VITROS Cl- or CREA Slides. Forward this notification if the product was distributed outside of your facility.
Initiated
2018-07-11
Root cause
Process control
Terminated
2020-08-19
Distribution
Worldwide distribution. US nationwide, Australia, Belgium, Bermuda, Brazil, Canada, Chile, China, Colombia, Denmark, France, Germany, India, Italy, Japan, Mexico, Norway, Poland, Portugal, Russia, Singapore, Spain, Sweden, The Netherlands, and United Kingdom.
Recall number
Z-3193-2018
Recall status
Terminated
CDRH record ID
166203
FDA record created
2018-09-19
Recall event number
80715

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗