Public FDA record

Z-3193-2026: Captus 4000e Thyroid Uptake System, Models: 1. 5430-30151, 2. 5430-30152, 3. 5430-30154;

Recall record Z-3193-2026 for Mirion Technologies (Capintec), Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-3193-2026
Company in FDA record
Mirion Technologies (Capintec), Inc.
FDA classification date
Sep 16, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Potential for failure of the tension bolt which would allow the spring arm to unexpectedly fall to its lowest position, which is approximately 25 inches (63.5 cm) from the ground.

Source facts

Action
On July 20, 2026 URGENT SAFETY NOTICE letters were emailed to customers. Actions to be taken: 1. Review this Alert and ensure that all affected personnel, including all operators of Captus 4000e Thyroid Uptake Systems, are aware of the contents. 2. Operate and inspect the device per the instructions provided. If your inspection shows that the collimator is not properly attached, please contact Capintec Technical Support at the phone number or email listed below. If you have any questions about this Inspection procedure of your Captus 4000 System, please contact Capintec Technical Support. If you notice any unusual change in the performance or functional response of the arm in your system (e.g. arm is making an unusual noise or arm does not move smoothly at any articulation joint), please contact Capintec Technical Support immediately. Capintec Technical Support: 1-800-631-3826 1-201-825-9500 capintecsupport@mirion.com
Initiated
2026-07-20
Root cause
Device Design
Distribution
Worldwide distribution - US Nationwide and the countries of ALGERIA, AUSTRIA, BANGLADESH, BRASIL, Brazil, CANADA, CHILE, Costa Rica, Ecuador, France, Hanoi Vietnam, Hong Kong, INDIA, ISRAEL, ITALY, Malaysia, Philippines, POLAND, Saudi Arabia, SOUTH KOREA, Spain, SWEDEN, TAIWAN, THAILAND, TURKEY, UNITED ARAB EMIATES, UNITED KINGDOM, VIETNAM.
Recall number
Z-3193-2026
Recall status
Open, Classified
CDRH record ID
222175
FDA classified
2026-09-16
Recall event number
99677

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗