Z-3194-2018: 2008T HEMODIALYSIS SYS, WITH CDX, Part Number 190713, Applicable only with 2008T BlueStar Premium Upgrade

Recall record Z-3194-2018 for Fresenius Medical Care Renal Therapies Group, LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-3194-2018
Company in FDA record
Fresenius Medical Care Renal Therapies Group, LLC
FDA record creation date
Sep 19, 2018
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

In the hemodialysis machine during the cleaning/disinfection program, when there is a power failure or interruption, the dialysis program button should be greyed out when the machine is powered up and a mandatory rinse should be required. However, if a power failure or interruption occurs during the cleaning/disinfection program with the auto-start feature enabled, the machine enters the dialysis program even though the dialysis program button is disabled and the mandatory rinse has not been performed. As a result, disinfecting chemical may not be completely removed from the machine before starting a treatment and disinfectant could exist in the dialysate delivered to the patient.

Source facts

Action
On 07/13/2018 phone calls were immediately conducted to the affected clinics. Contact was made to the Technical Service Directors explaining issue and to disable the Auto Start Function on their BlueStar SW v2.71 installed 2008T Hemodialysis Machines.
Initiated
2018-07-13
Root cause
Under Investigation by firm
Terminated
2019-09-17
Distribution
US Nationwide in the states of : CA, CT, MI, NC, and NY.
Recall number
Z-3194-2018
Recall status
Terminated
CDRH record ID
166788
FDA record created
2018-09-19
Recall event number
80875

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗