Z-3194-2026: Halyard IV START KIT NON-STERILE; 48 kits per case. Containing BD ChloraPrep Clear - 1 mL Applicator. Kit

Recall record Z-3194-2026 for AVID Medical, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-3194-2026
Company in FDA record
AVID Medical, Inc.
FDA classification date
Sep 23, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

Convenience kits containing BD ChloraPrep Clear - 1 mL Applicator, which may be contaminated with Aspergillus penicillioides.

Source facts

Action
Consignees were initially notified on about 07/22/2026 via "URGENT: MEDICAL DEVICE FIELD CORRECTION" letter. They were informed that product may exhibit an open or incomplete seal on the packaging of the applicator. Consignees were instructed to quarantine impacted product, discontinue use of the BD ChloraPrep Clear - 1 mL Applicators within the kits, add a warning label to instruct users to discard the impacted component, and notify customers if the kits had been further distributed. Consignees were notified again on about 08/20/2026 with an updated "URGENT: MEDICAL DEVICE FIELD CORRECTION" letter. Consignees were informed that certain lots of the BD ChloraPrep Applicators are contaminated with Aspergillus penicillioides. Instructions included to discard impacted kits containing affected applicators and notify customers if the kits had been further distributed.
Initiated
2026-07-22
Root cause
Packaging process control
Distribution
US Nationwide distribution.
Recall number
Z-3194-2026
Recall status
Open, Classified
CDRH record ID
222426
FDA classified
2026-09-23
Recall event number
99761

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗